This manual provides a basic operation of Varian NMR spectrometers with Vnmrj 2.2D and above and Chempack 4.1. It also describes the instruments' capabilities and limitations, and provides information on how to reserve instrument time and access spectra stored in the spectrometers. The manual also covers Vnmrj overview, including its modes of operation and the concept of workspaces.

Typology: Study notes

2022/2023

Uploaded on 05/11/2023

aghanashin
aghanashin 🇺🇸

4.7

(22)

253 documents

1 / 21

Toggle sidebar

This page cannot be seen from the preview

Don't miss anything!

bg1
Vnmrj Training Guide
Eugenio Alvarado.
Department of Chemistry, University of Michigan. 10/29/10, 05/05/11, 07/02/14
This manual describes the most basic operation of our Varian NMR spectrometers with
Vnmrj 2.2D and above and Chempack 4.1. More advanced techniques, like two-
dimensional spectroscopy, are described in separate documents in the NMR Facility's
web site. Full documentation for Vnmrj is available from its help menu and on printed
copies of the manuals located the lab.
Please note that reading all the information in the Training page of our web site is a
requirement before you start using our Facility. You are also required to read and
understand our instrument's capabilities and limitations. This information can be found in
the Instruments page of our web site. There is also a training video available in one of
our workstations (Krypton).
To reserve instrument time, we use an online scheduler which is accessed from our web
site at the bottom of the News page. On that page, we regularly publish important news
and information about the NMR Facility. Please check the news every time you want to
make a new reservation for current news and information.
Currently, the lab is equipped with five Varian NMR spectrometers: two 400MR (named
Gallium and Cobalt), one Inova 500 (Tin), one VNMRS 500 (Tellurium) and one VNMRS
700 (Ytterbium). These have a variety of capabilities and each is best suited for some
particular experiments. For example, Cobalt has better sensitivity for carbon-13 than Tin,
but the latter has better sensitivity for protons. For a complete list of capabilities and
limitations, consult our web site or ask the staff.
The lab also has five workstations where users can process and plot their spectra. One
of them (copper) can also operate with the Windows operating system and has available
a copy of MestReLab's MestReNova for NMR processing and some other chemistry-
related software. Spectra stored in the spectrometers can be accessed from the
workstations via a local network. Use of the workstations is free of charge.
If you encounter a problem when using Vnmrj or one of the spectrometers, read
Common Problems and Solutions” in the Documentation section of our web site or ask
the staff for help.
Vnmrj overview
Vnmrj runs on computers under the Linux operating system. Users in our facility receive
an account on each computer in the lab which can be accessed with their UM uniqname
(user name) and password. Each user has a home directory or folder located in
/home/username. Inside that folder, Vnmrj stores its user-related files in a folder
named vnmrsys and data files (spectra) in vnmrsys/data
(/home/username/vnmrsys/data).
Vnmrj can operate in three modes: experimental, walkup and imaging. Because we use
only the experimental interface by default in our Facility, this manual will cover only that
interface. The walkup interface is a simplified mode of operation with many limitations
and is meant mostly for automation.
pf3
pf4
pf5
pf8
pf9
pfa
pfd
pfe
pff
pf12
pf13
pf14
pf15

Partial preview of the text

Download Vnmrj Training Guide and more Study notes Chemistry in PDF only on Docsity!

Vnmrj Training Guide

Eugenio Alvarado. Department of Chemistry, University of Michigan. 10/29/10, 05/05/11, 07/02/ This manual describes the most basic operation of our Varian NMR spectrometers with Vnmrj 2.2D and above and Chempack 4.1. More advanced techniques, like two- dimensional spectroscopy, are described in separate documents in the NMR Facility's web site. Full documentation for Vnmrj is available from its help menu and on printed copies of the manuals located the lab. Please note that reading all the information in the Training page of our web site is a requirement before you start using our Facility. You are also required to read and understand our instrument's capabilities and limitations. This information can be found in the Instruments page of our web site. There is also a training video available in one of our workstations ( Krypton ). To reserve instrument time, we use an online scheduler which is accessed from our web site at the bottom of the News page. On that page, we regularly publish important news and information about the NMR Facility. Please check the news every time you want to make a new reservation for current news and information. Currently, the lab is equipped with five Varian NMR spectrometers: two 400MR (named Gallium and Cobalt ), one Inova 500 ( Tin ), one VNMRS 500 ( Tellurium ) and one VNMRS 700 ( Ytterbium ). These have a variety of capabilities and each is best suited for some particular experiments. For example, Cobalt has better sensitivity for carbon-13 than Tin , but the latter has better sensitivity for protons. For a complete list of capabilities and limitations, consult our web site or ask the staff. The lab also has five workstations where users can process and plot their spectra. One of them ( copper ) can also operate with the Windows operating system and has available a copy of MestReLab's MestReNova for NMR processing and some other chemistry- related software. Spectra stored in the spectrometers can be accessed from the workstations via a local network. Use of the workstations is free of charge. If you encounter a problem when using Vnmrj or one of the spectrometers, read “ Common Problems and Solutions ” in the Documentation section of our web site or ask the staff for help.

Vnmrj overview

Vnmrj runs on computers under the Linux operating system. Users in our facility receive an account on each computer in the lab which can be accessed with their UM uniqname (user name) and password. Each user has a home directory or folder located in /home/ username. Inside that folder, Vnmrj stores its user-related files in a folder named vnmrsys and data files (spectra) in vnmrsys/data (/home/ username /vnmrsys/data). Vnmrj can operate in three modes: experimental, walkup and imaging. Because we use only the experimental interface by default in our Facility, this manual will cover only that interface. The walkup interface is a simplified mode of operation with many limitations and is meant mostly for automation.

A data set in Vnmrj consists of the NMR data itself (FID) and its associated parameters for acquisition and processing; and it is stored not as a single file but as a folder with an .fid extension containing several files. Data sets are loaded from the File menu. In addition, Vnmrj uses the concept of “ workspaces ” (analogous to opening several files in a word processing program) to perform several tasks simultaneously. Workspaces (also called experiments ) are separate regions of computer memory independent of each other where spectra and parameters are kept. For example, you can be performing a proton acquisition in workspace 1, while setting up acquisition parameters for a carbon-13 in workspace 2, or plotting a two-dimensional spectrum in workspace 3. The number of available workspaces is essentially unlimited, but they must be created once before they can be used. Only workspace 1 already exists in any user account by default. Once created, data in any workspace is persistent between user logins. To create an additional workspace, select from the main menu: File > Create Workspace. To move between workspaces, type a command jexp (“ j oin exp eriment”) immediately followed by a number; for example jexp3 will take you to workspace 3, if it already exists. Or you can also select one from the menu Tools > Workspace Information.... The workspace you are working on is displayed in the blue bar on top of the graphics display. The dynamic use of workspaces increases the efficiency when using Vnmrj and is highly recommended. For example, after acquiring a proton spectrum in experiment 1, you can copy the FID to experiment 3 with the command mf(1,3) (move fid from 1 to 3) and setup a 2D experiment based on the parameters you selected for the 1D spectrum. Please refer to the figure below for a brief description of the layout of the graphical interface. Vnmrj experimental interface: Main menu User tool bar Vertical panels Command line Command output Vertical tool bar Graphics display Parameter panels tabs Action buttons Parameter panels Hardware bar Status of temperature, spin and lock Current probe Probe setup System status Message window Current workspace Pulse sequence Array element

Do not hold or touch the spinner on the shaft Do not hold the spinner from the NMR tube. Do not insert the tube by forcing it in the spinner's hole, or when the spinner is mounted in the depth gauge. If less than the recommended volume is used, center the solution's column with the center black line on the gauge's meter.

If you break your tube and spill the solution, you are responsible to

thoroughly clean everything up and make sure no hazard remains to

other users or the instruments. You should also let us know

immediately to make sure everything is satisfactory. Users who

don't show consideration for fellow users or our instruments will be

banished from the NMR Lab.

Preparing for acquisition

If you absolutely don't need more than simple, routine spectra; don't want to learn to optimize your experiments; or don't even care about the details of NMR spectroscopy but need some NMR spectra, there is an automation mode that can do everything for you except for loading the sample in the magnet. If that is what you want, go to the section Automated Acquisitions ; otherwise keep reading.

Before you can run any NMR experiment you will need to specify a pulse sequence to be executed and several acquisition parameters like pulse durations, frequencies and powers; delays; the number of data points; and many others. Additionally, many of these parameters require proper calibration. All these parameters have been calibrated and given default values by the staff, and are stored on files in the computer's hard disk. To run an NMR experiment, you must first load a set of parameters. The parameters are then kept in portions of memory called workspaces where acquisition and processing will take place. The value of the parameters can be changed on the command line by typing the parameter's name followed by the equal sign and the new value, for example sw=5000. If the value is alphanumeric, it must be enclosed in single quotation marks, for example gain='n'. To find the current value of a parameter, type its name followed by a question mark, for example nt?. You can also inspect and change most common parameters from the appropriate parameter panel. To load a parameter set, select an experiment from the Experiments menu. Notice that the Experiments menu is divided in 3 parts. The sections named Setup NEW parameters for... and Setup NEW parameters to do... will replace all current parameters with default values for the experiment requested; while Convert current parameters to do... will use the existing parameters in the current workspace and add any additional parameters needed for the experiment. For example, before doing a 2D experiment, it is a good practice to first measure a 1D proton spectrum, check on the quality of the shimming, and optimize parameters (e.g. spectral width, transmitter offset, relaxation delay). To do this, first select Setup NEW parameters for... > Proton , modify the parameters, take a spectrum and save it; and then select the 2D experiment from Convert current parameters to do.... But if you don't care to optimize anything and just want your 2D spectrum with default parameters, you can choose it directly from Setup NEW parameters to do.... After the sample has been inserted but before the acquisition is started, the spectrometer must be prepared. This involves enabling the lock, tuning the probe and shimming the magnet. First, go to the Start, Standard parameter panel and choose a solvent. At this point you can enter a Comment to identify your sample, and choose a different spinning rate or temperature. We will deal with variable temperature operation later. The information in Sample Name , Lot Number , Notebook , Page and Email is used by vnmrj's database to locate spectra and is optional for normal operation.

have vnmrj find it for you, but be aware that Find z0 may reset the lock power to a saturating value. Z0 is far from resonance Getting closer...^ Z0 is on resonance The lock power controls the RF radiation's power that is applied to the solvent's deuterium signal. Low powers give very weak signals and a poor, noisy lock while too high powers produce signal saturation leading to lock instability. The optimum power is determined mainly by the relaxation time of the solvent and probe design. For example, in the Inova 400 the optimum power for CDCl 3 is about 24 and that for acetone-d 6 is about 6; using a power of 24 for acetone will certainly lead to severe saturation and instability, and shimming will be impossible. The lock gain provides amplification to the lock signal. Its value is not really important as long as a lock intensity of at least 60% is maintained during acquisitions. The lock phase is mainly determined by the electronics and to a lesser extent by the solvent. Usually it is not critical to optimize it (the default parameters already have a good value) as long as an intense lock is produced. However, if you are unable to lock, make large changes (of ~30 units) and see if it helps. Optimize it if needed. Clicking the On button in the Start, Lock panel engages the lock system. If the lock system didn't engage (the lock level is less than 10%), try increasing temporarily the power or the gain to kick it in, but once the lock is set, turn the power back to its previous value or you may saturate the lock signal.

Tuning the probe

In a typical NMR experiment, the nuclear spins in the sample are excited by the application of one or more short pulses of radio-frequency (RF) radiation, and eventually they precess freely across the axis of the external magnetic field, thus inducing an electric current in a coil located close by. The coil is located in the probe, whose main purposes are to hold the coil(s) close to the sample and to generate the strongest possible signal from the weak fluctuating currents induced by the nuclear spins as they precess. For the coils to be able to transmit the RF pulse to the sample and to detect the induced NMR signal efficiently, the coil circuitry should be optimized. The process is known as “tuning”, and it is analogous to tuning your radio receiver to a particular radio station: if not properly tuned, your music will be noisy. Proper tuning is essential to obtain high quality NMR data. The exact frequency depends of course on the nucleus to be measured, but also on the solvent. The adjustments are made by turning the tuning and matching rods located on the bottom of the probe, which in turn adjust variable capacitors inside the probe body. In our lab, the procedures for tuning are different depending on the probe used:

  • The probes in the Inova-400 and in Cobalt are permanently tuned to 1 H, 19 F, 13 C and 31 P and do not require additional tuning.
  • The default probe in the Inova-500 is always tuned to 1 H and 13 C and does not require further tuning by the user but can be tuned upon request for special

experiments. This instrument has another probe that can be tuned to any nucleus from 31 P to 15 N. If needed, the staff will install and tune it; users are not allowed to do this except when properly trained.

  • The other MR400 instrument ( Gallium ) is equipped with an automatic tuning device (protune unit) that can tune the probe for the user. The software is set up to do automatic tuning before acquisition when a new sample is inserted in the magnet or when the nucleus or the solvent is changed. All you have to do is start the acquisition by clicking the Acquire button (typing go or ga will not do the tuning). Alternatively, you can request the tuning to be done at any time by clicking the Tune button on the Start, Standard panel and then click the appropriate button for your nucleus; or type in any desired nucleus (letters first, numbers last: eg. Pb207) followed by clicking Tune to Criterion.
  • The VNMRS instruments ( Tellurium and Ytterbium ) are also equipped with protune units that can tune the probe for the user as explained above. Additionally, these instruments require a “quarter wavelength cable”, matched to the frequency of the nucleus, to be changed at the preamp interface before acquisition of heteronuclei other than 13 C. Ask the staff for details. Unfortunately, the protune unit is notoriously unreliable. It usually does a good job, especially for 1 H, 19 F and 13 C, but for other nuclei it may fail. This is especially true for some nuclei like 11 B and 199 Hg in our 400-MR spectrometers Cobalt and Gallium. If automatic tuning fails (characterized by the motors in the unit moving up and down for more than 1 minute), you will have to tune the probe manually as described in a separate writeup ( Probe tuning with the protune modules ). For this reason it is always a good idea to verify if tuning was successful, especially before starting a long acquisition and it will only take a few seconds. From the main menu choose Tools > Probe Tuning > Manually Tune Probe. Then select Tune RF channel (1 for H1 or F19; 2 for anything else) and enter a nucleus in the Center Frequency field (letters first, numbers last: eg. Pb207). After clicking on Start Probe Tune the display should show a V-shaped dip centered with the blue line; if not, tuning was not successful and you will need to manually tune the probe or try automatic tuning again. Click Stop Probe Tune followed by Quit to return to normal operation.

Shimming

One of the most important factors to get a good quality NMR spectrum is shimming. NMR spectroscopy demands a homogeneous magnetic field over the whole of the sample volume located within the probe's detection coil. If the field is not homogeneous,

higher the order of the shim, the closer to the peak's base its effect will be. Large spinning side bands indicate poor non-spinning shims. These are shims with components in the X or Y axes and as the name implies, they must be adjusted with the spinning turned off. If you observe large spinning side bands, turn the spinning off and adjust X and Y interactively using the procedure for Z1-Z2. Keep in mind that some solvents, particularly acetone and methanol, react more slowly to changes in the shims due to their typically long relaxation times. With these solvents, shimming has to be done with low lock powers, slowly and patiently. You can also use automatic gradient shimming on all our instruments. This procedure shims Z1 to Z5 simultaneously and usually achieves satisfactory results. To use it, click the Gradient Shim button in the Start, Standard panel and wait until it finishes. If your tube requires more than the normal shimming, it may be necessary to run it twice. Give it a try, it may even give better results than your manual shimming! At the very least, it will provide a much better starting point to start manual shimming and doesn't even require the sample to be locked. Also, if you are unable to find the lock signal, try gradient shimming first. It is also possible to shim non deuterated solvents. To do it, select the equivalent deuterated solvent from the list, turn the lock off and select from the main menu: Acquisition > Go Gradient Shimming > Use 1H/gradient map. Locking is not possible, of course, but this is not a problem with short term acquisitions, like routine 1D spectra, as the magnetic field in modern superconducting magnets is very stable. Gradient shimming of non deuterated solvents, and manual shimming using the FID is described in more detail in a separate writeup. A special situation may arise when acquiring a fluorine spectrum in a non deuterated solvent. On spectrometers with automatic tuning capability, it is important to first take a proton spectrum to ensure the probe is tuned to proton, then do gradient shimming with the proton gradient map and then load parameters for fluorine and tune the probe to fluorine. If you don't follow this order but start by taking a fluorine spectrum, automatic shimming will fail because the probe will be tuned to fluorine before shimming with the proton gradient map is performed.

Variable temperature operation

For temperatures above room temperature, all you need to do is enter a value in the Start, Spin/Temp panel and click Regulate Temp. For low temperature operation we have two choices. All spectrometers are equipped with an FTS cooling unit that can be used to cool the probe to temperatures up to about -20 °C. For lower temperatures, a bucket filled with liquid nitrogen is used. Ask the staff for further instructions before using low temperature for the first time. In either case, be careful no to exceed the limits of the installed probe or the boiling or freezing points of your solvent. Make incremental changes to the temperature not larger than 30 °C until you reach the desired temperature. Keep in mind that the sample will require at least 5 to 10 minutes to fully equilibrate to the temperature in the probe, and you may need to re-shim or re-tune after large changes in temperature. When you are finished, bring the temperature back to room temperature, again in steps of 30 °C, and finally click on Temp Off in the Spin/Temp panel to turn off the temperature controller. Allow enough time after your experiments for the temperature to recover completely before your time is up. The next user should find the probe at room temperature!

Modifying spectral parameters

Before starting the acquisition, you may want to adjust the most common spectral parameters. These are found in the Acquire, Defaults parameter panel. Select a different Spectral Width from the list or type in the chemical shift range. The default values for some nuclei can be very far from optimum for your sample or may not even encompass the values you expect. Narrowing the range will increase the digital resolution. Select a Number of scans (number of transients: parameter nt) or type any number you want in the command line. For example, low sensitivity nuclei or diluted samples may require a large unknown number of scans. In such cases, enter a large number like nt=100000 and stop the acquisition later when enough signal has been accumulated. To get a more accurate integration you may need to increase the Relaxation Delay (parameter d1). A Pulse Angle of 45 or 60 degrees is usually adequate for protons, but in order to get quaternary carbons in a carbon-13 spectrum it may help to lower the pulse angle to 30 degrees. Other common parameters may be changed from the Acquisition panel or typed directly. For example, the acquisition time (at) determines the digital resolution of the spectrum; if you need high resolution fluorine spectra to accurately measure coupling constants, you should reduce the default spectral width and increase the default acquisition time. Do not change parameters that you don't understand and never increase pulse or decoupling powers. When doing experiments other than simple 1D spectra, it is important, and sometimes necessary, to review and perhaps change the parameters in the Pulse Sequence panel, where the parameters pertinent to the chosen pulse sequence are accessible. Consult the experiment's documentation. A note about spinning. Although the default parameters for common nuclei set a spinning rate of 20 ( Start, Standard or Spin/Temp panels), this is not really necessary. The resolution in a well shimmed, modern superconducting magnet is usually such that you may not even notice a difference in the line widths with and without spinning, but you may be able to see spinning side bands when the sample is spinning and the non-axial shims are far from optimum. In fact, most 2D spectra are recorded with spinning off. Sometimes, the sample will not spin because of dirt accumulated inside the magnet. Please let us know if this happens so we can clean the magnet but don't worry about your spectrum and just turn spinning off. Before starting the acquisition, you can check the time required by the experiment by clicking Show Time. Click Acquire or type go, or ga to start the acquisition. After it has started, it can be stopped at any time by clicking Stop or typing aa (abort acquisition).

Arrayed acquisitions

Some experiments, like a kinetics or a diffusion experiment, actually consist of collecting not one, but a series of spectra. The set is acquired as a single experiment and stored as a single file. To perform the acquisition, one of the parameter's is set up as an array of values. Vnmrj then acquires the first spectrum with the parameter set to the first value in the array, then the second spectrum is collected with the parameter set to the second value and so on. For example, to quickly determine the 90 degree pulse duration, you can set up an array of pw values (the pulse width, in microseconds). You can do it by typing the parameter followed by the values separated by an equal sign and commas, with no spaces: pw=2,4,6,8,10,12,14; or you can setup the array from the menu Acquisition > Parameter Arrays... The instructions for the experiment you want to run will normally tell you what parameters to array and how.

Automated Acquisitions

If you want a completely automated acquisition, proceed as follows:

1. Insert the sample in the magnet as usual. 2. Click New Study in the lower part of the Holding vertical panel (shown below), or select from the menu Automation > New Automation. 3. Enter a Sample Name in the Start, Standard panel, and select the solvent and other options. You can select Autoplot to have each spectrum plotted; autolock ( Lock? Yes ) to capture the lock automatically, Shim to shim the sample and Tune to tune the probe if you haven't done it yourself. 4. In the Experiment Panel , click the buttons for the desired experiments (some, like noesy1D require further customization). 5. Click Submit. The experiments selected in step 4 will have default acquisition parameters that may or may not be suitable for you. Typically, you may need to change at least the number of transients or the spectral width. The experiments may be customized by double clicking the experiment name (in green) in the Study Queue window (not the experiment button). The parameter panels for that experiment will open up to allow you to make modifications as usual. Click Save to save any modified parameter and return to the

queue entry panel. Also on the Study Queue window, notice the acquisition times for the individual experiments displayed inside brackets to the right of each experiment. The total time for the queue is displayed at the bottom of the Start, Standard panel. After each acquisition, the spectrum will be processed and plotted automatically (if requested) and the group of experiments (the queue) will be saved in a folder with a name consisting of the sample name, followed by the date and a sequence number separated by underlines; for example: ealva-1-234-2_20101026_01. Each FID will be saved inside that folder with a name consisting of the sample name followed by the experiment name followed by a sequence number; for example: ealva-1-234- 2_HMQC_01.fid. If you need expansions of your spectra, you will have to open each spectrum and plot them manually yourself.

Processing

When you save a spectrum, the file is stored in the spectrometer's computer, usually in your vnmrj folder in /home/ uniqname /vnmrsys/data. Your files will be accessible from any workstation in the lab. To load a spectrum stored on disk into the current workspace, select Open... from the File menu, or click the Open... icon in the user tool bar just below the main menu (hover the mouse cursor over the icons to display their function). When you use this utility on any workstation in the lab, you will see links to all spectrometers. Clicking a link allows you to browse the files and load an FID stored in that instrument. If Process data on drag-and-drop is enabled ( Edit > System settings menu), then the data will be Fourier transformed and processed automatically for you and it will be almost ready to plot: you may still need to check the referencing and phase as described below. If automatic processing is not enabled, you can do automatic processing from the menu Process > Full Process. Processing an NMR spectrum generally involves applying a weighing function to the FID; applying linear prediction if needed; doing a Fourier transformation; phase correction; adjusting the chemical shift reference; selecting integral regions; baseline correction if needed; and possibly getting a list of peak positions. Most of these task are done during automatic processing. However, it is important to understand how to do them manually because the automatic algorithms may not be completely accurate for all spectra. The parameters for processing can be adjusted from the Process panels. Many of these parameters can be found in more than one panel and also in some of the vertical panels on the left hand side of vnmrj. The repetition is meant for ease of use but sometimes can be confusing. If you are not sure about a parameter or its value, don't change it.

Weighting function

Select a weighting function and its parameter(s) on the Weighting panel to be applied before Fourier transformation. The parameter for zero-filling is also found here.

Linear prediction

Linear prediction is also applied before Fourier transformation. Its parameters are located in the More 1D or More 2D panels. There is more information on linear prediction in a separate writeup.

any movement in any direction above the point where you clicked stretches the spectrum, and to compress it you have to move the mouse to a position below the point you initially clicked, possibly stretching it even more as you move the mouse down. Click the top button in the toolbar to exit this mode.

Phase correction

The Free Induction Decay (FID) data set is a function of a complex number, the intensity, versus time. During Fourier transformation, this function is converted to a function of intensity (but still a complex number) versus frequency. The spectrum displayed on the screen should correspond to the real or absorptive portion of the function, but it may contain “imaginary” or dispersive contributions as shown in the spectrum below. If you chose automatic processing or automatic phase correction ( Autophase button in Process, Default panel ), the spectrum will probably be phased correctly. But on occasions, you will need to fine adjust it manually. To phase a spectrum, first click the Phase button to enter into phase correction mode. Left click on any portion of the spectrum close to its right edge and drag the mouse up or down until the rightmost peaks in the spectrum have only positive components; don't worry about the peaks on the left side of the spectrum yet. The right mouse button can also be used for fine adjustments. Then, left click close to the left edge of the spectrum and drag the mouse up or down until the peaks on the left side of the spectrum are completely in phase (positive intensity). Repeat the procedure if necessary until the full spectrum is properly phased. Click the top button on the toolbar to get out of phase correction mode.

Chemical shift referencing

Generally, the chemical shifts in the spectrum should be acceptable. By default, the chemical shifts are referenced using the solvent peak's deuterium signal. However, if you want to be more accurate, or if you have an unusual or non deuterated solvent, you may need to reference the chemical shifts manually using an internal standard like TMS. The reference can be set on the Process, Default or Display panels. Expand a small region around your internal standard, place the cursor in its center and enter the standard's chemical shift in the Reference cursor to field. After acquisition, the chemical shift of any

other nucleus will also be automatically referenced using the solvent's deuterium peak. But IUPAC recommends using the proton peak of the internal standards TMS for organic solvents and DSS for H 2 O or D 2 O. To reference any other nucleus in this manner, proceed as follows:

1. take a proton spectrum in workspace 1 and set the chemical shift of the internal standard (0 ppm for TMS or DSS) 2. take the spectrum of the nucleus of interest in another workspace, for example 2 3. type mref(1,2) in the command line to “move the reference” from workspace 1 to workspace 2. Your spectrum is now referenced according to IUPAC without the need for cumbersome external references! ( Pure Appl. Chem. 2001;73:1795-1818).

Integrating the spectrum

As a part of autoprocessing, integral regions are selected and the integrals' level and tilt are corrected. You may only need to normalize the areas by assigning a value to one of the integrals. To do it, first click the button in the vertical toolbar to display the integrals; place the cursor over one of the integrals, and in the Process, Integration panel select Normalize Area To: Single Peak ; enter a value in the Integral Area field and click the Set Integral Value button. The integrals' values can be displayed with the button Display List of Integrals or with Show Integral Values. If you want to select integral regions manually, first reset the regions previously defined, if any, with the Clear Integrals button. Then, select Partial under Integral Display Mode , click on Interactive Resets and define the regions by clicking on the spectrum over the left and right limits of each region that you want to create. Always define the areas so they enclose some empty baseline on both sides. When you are done, click the top button on the vertical toolbar to get out of this “region selection” mode. It is also common to find that the whole integral “baseline” is not straight and horizontal: it may be tilted and may show a curvature. If that is the case, to get accurate integrations you will need to correct it manually. Select Full under Integral Display Mode and click the Integral Lvl/Tlt button in the toolbar. Next, click on the left side of the spectrum and drag the mouse up and down until the integral baseline on that side of the spectrum is horizontal, then click on the right side of the spectrum and drag the mouse until the integral baseline is straight and horizontal across the spectrum. You may need to repeat the procedure to obtain optimum results. As usual, click the top toolbar button to get out of this operation mode. Notice that it is common to see some curvature or tilt

component is selected by pushing a button on the Process, Plot panel shown below. First, choose the plotting options for the spectrum, parameters, integrals and peak frequencies. These are selected from the drop down menus in the Plot Setup section of the panel. For the Plot (spectrum) you usually want to select As Displayed to plot only the region currently displayed on the screen. For the parameters, Basic (left) prints only the most important parameters, and Full (left) prints almost all acquisition and processing parameters. To plot the integrals, they must first be displayed on screen, and Scaled, Horiz. is the most common option. You can choose any of the options and click on Show to preview your selection on the screen. For the peak frequencies, there are quite a few choices suitable to different spectra; the most common are On Peaks, top, ppm and As a List. Other options, like the width or height of the plot can be controlled from the command line or from the ArrayedSpectra vertical panel. Once the layout is chosen, the easiest way to plot, as usual, is to select an automatic procedure. You can either click Automatic Plot Page to have a plot with the spectrum, scale and parameters sent directly to the printer, or click Auto Plot Preview to send it to an Adobe Acrobat Reader window as a pdf image for your review. Or you can create the plot manually by clicking the elements that you want, like Plot Spectrum , Plot Spectrum Scale , Plot Parameters , Plot Integrals , etc. Once you clicked the elements, click on Plot Page or Plot Preview to send your composition to the printer or to Acrobat Reader respectively. Clear Plot will remove your composition from memory to start all over; and our custom macro, UMmailpage, will send you the plot via email as an encapsulated postscript file and a high resolution image file (png format) ready to be inserted into word processing or presentation documents. Although our laser printer prints in black and white, plotting in color is possible if you create an image or plot file as described below and print it on your own color printer. The options or colors are selected in the Styles and Themes window which is displayed when you select Display options.. from the Edit menu. Two themes have been preconfigured in our lab: UM_color and UM_gray. The first plots the elements of the plot in different colors, and the second uses only black and different shades of gray. Choose a theme on the drop down menu of that window and click Load. The colors can also be changed, if needed as follows. For 1D plots, select the DisplayColors tab and Plot Postscript in the drop down menu. Then click on the color button or on the drop down menu to the right of each element and select or create a color. For 2D spectra, select the 2DColors tab and Plot Postscript from the drop down menu. Then click on either Absolute or Phase for the kind of 2D spectrum to plot

(for example, a cosy or an hmbc are usually displayed in absolute mode and an hsqc or noesy in phase sensitive mode). Select a color for each of the contour levels on the plot as described above for the 1D elements. Click Close to accept the changes or Abandon to discard them and finish. To plot with a different line width, create the plot manually as described above and then type pagelw( x ), where x is the desired linewidth in points. The default width is 3: choose 1 for the finest and 5 to 9 for thicker lines.

Miscellaneous

There is more information about doing other common tasks, like transferring files via ftp or flash drives, or burning DVDs in the Frequently Asked Questions writeup available in our web page.

Recovering from problems

Problems arising from the hardware or software will inevitably occur. The most common problems and their solutions have been compiled in the document Common Problems and Solutions which is available in our web site and also in a white binder in the lab. Of these, possibly the most common is due to disrupted communications between the host computer and the spectrometer's console. This one can manifest itself as an unresponsive vnmrj or in many other ways, and is usually fixed by typing abortallacqs in the command line. Please consult Common Problems and Solutions for details. Never power cycle the computers. You will risk a real probability of corrupting the file system beyond recovery.

Our instruments

Currently we have six Varian (now Agilent) spectrometers in our facility. Additional technical details can be found in our web page as well as current charge rates.. Gallium is a 400MR spectrometer equipped with an AutoX-DB direct-detection probe and automatic tuning. The temperature range is -80 to +130 ºC. This probe is capable of detecting 1 H, 19 F and any nucleus with a frequency between those of 31 P and 15 N. It is more efficient for X-nuclei detection experiments (nuclei other than 1 H or 19 F). Cobalt is a 400MR spectrometer equipped with a auto-switchable probe that is permanently tuned to 1 H, 19 F, 13 C and 31 P. The temperature range is -20 to +80 ºC. This instrument is currently in walk-on mode of operation (no previous reservation is required) and is meant for quick routine analyses of synthesis products. Tin is an Inova 500 spectrometer. It has two probes available: the id probe (the default) is an indirect detection probe without fluorine capability, temperature range -20 to 50 ºC optimized for proton detection; the sw probe is an indirect detection probe, temperature range -100 to 150 ºC, optimized for heteronuclear detection. The probes are switched and manually tuned by the staff upon request. Zirconium is an Inova 400 spectrometer. It has only one probe that is permanently tuned to 1 H, 19 F, 13 C and 31 P. The temperature range is -20 to 80 ºC. It is currently installed in room 2407 and is used for undergraduate teaching. Tellurium is a VNMRS 500 spectrometer with automatic tuning, three RF channels and two probes. The OneProbe (the default) has a temperature range of -80 to 130 ºC and excellent sensitivity for 1 H, 19 F, 13 C and other heteronuclei and is suitable for all kinds of experiments. The HFX probe is tuned to 1 H, 19 F and X, with a temperature range of -